RX setup

N˚ of low contrast inserts detected

N˚ of small detail inserts detected

Spatial resolution detected [lp/mm]

S 100

S 200

S 100

S 200

S 100

S 200

55 kV 3.2 mAs

11

12

14

13

3.15

2.24

55 kV 3.2 mAs 4 mmAl

9

7

11

9

2.5

2.0

60 kV 1 mAs 4 mmAl

7

7

10

10

2.5

1.8

60 kV 1 mAs

11

9

12

11

2.8

1.8

62 kV 0.4 mAs

9

6

9

10

2.5

1.6

60 kV 0.4 mAs

8

6

9

6

2.24

1.4

60 kV 0.32 mAs

7

4

8

1

2.24

1.25

60 kV 0.2 mAs

7

2

7

0

1.4

0.9